采用5.7英寸(120×90)PLC多彩液晶觸摸顯示屏,中文菜單顯示。公稱規格、管壁、測試次數、加壓、上行、下行、時間、標定。由鍵盤控制液晶顯示屏上的菜單,機載打印測試數據。實時顯示刺穿力和峰值力數據,可顯示刺穿力及F五種峰值力。CL15811-D科研注射針針尖刺穿力測試儀*符合GB15811《注射針》標準中相關條款設計制造。
It uses 5.7 inch (120 *90) PLC multi-color LCD touch screen to display Chinese menu. Nominal specification, pipe wall, test times, pressure, up and down, time and calibration. The menu on LCD screen is controlled by keyboard, and the test data is printed on board. Real-time display of piercing force and peak force data can show the maximum piercing force and F five peak forces. CL15811-D Medical Injection Needle Point Puncture Tester fully complies with the relevant provisions of GB15811 Standard for Disposable Sterile Injection Needles.
公稱規格:0.3~1.2mm(0.2~3.4mm);其它規格:0.12-4mm可選,銷售
測試范圍:0.10~5.00N、精度:±0.01N;
移動速度:100mm/min;
(贈送)模擬皮膚:聚氨酯膜材料,符合GB15811標準;
打印方式:每次可打印8次、13次或20次測試數據及平均穿刺力和偏差值。
凈重:11kg 外形尺寸:400×290×240(mm)
Nominal specification: 0.3-1.2 mm (0.2-3.4 mm);
Test range: 0.10-5.00N, accuracy: +0.01N;
Moving speed: 100mm/min;
(Gift) Simulated skin: polyurethane membrane material, in line with the GB15811 standard;
Printing method: 8 times, 13 times or 20 times test data and average puncture force and deviation value can be printed each time.
Net weight: 11kg shape size: 400 x 290 x 240 (mm)
注意事項
6.1 操作人員必須熟悉本測試儀使用說明書。
6.2 儀表中設立一檔快速轉跳鍵。
當光標在‘測試’鍵以下時,按‘←’鍵,光標將跳轉到‘測試’。
6.3 開機后儀表進行自檢測,發現錯誤,在第1屏結束后將出現提示屏,提醒你重新設定測試方案或重新標定。
出現‘Ⅱ類數據錯誤’,你應該重新設定測試方案。
出現‘Ⅰ類數據錯誤’,你應該請專業人員重新設定標定。
6.4 本測試儀定期計量周期為一年。
6.5 本測試儀實行三包,三包期限為一年。
6.6 非專業人員不得進入‘標定’!