產品簡介
全模擬訊號輸出的ECG數據庫播放器,協助您完成系統及算法驗證,適用于研發及法規符合性測試??芍苯蛹虞d及撥放各類型心電數據庫,符合國際及國內新版標準:IEC 60601-2-25、IEC 60601-2-47、ANSI/AAMI EC57、YY0782、YY0885
產品特點
可供心電診斷、可攜式心電設備仿真數據庫產生器
支持12 導聯,8 個信道同步產生數據庫訊號
內建CAL、ANE 及Biological 波形
可直接加載及播放AHA、MIT、CU 數據庫,無須格式轉換
較佳的輸出訊號準確度,電壓輸出<±8μV
提供軟件開發工具包(Software Development Kit) 讓用戶自行開發全自動化測試軟件
產品規格
Item | Details / Reference | Value |
---|---|---|
Output channels | The 8 output channels are provided through a network as specified in IEC 60601-2-51 to provide signals to 10 lead electrodes; in the device under test, this will be displayed as 12 leads. | 8 outputs 10 lead electrodes 12 leads |
Voltage accuracy | IEC 60601-2-51 specifies a limit of ±1%, but does not provide a lower limit (all systems must have a lower limit). An inferred specification of 1% ±5µV is derived from the device under test specification in IEC 60601-2-51 of 5% ±25µV. | ±1% for |
Output voltage | The output voltage on most of the database and ECG is +5mV~-5mV. | ±5mV |
Output noise level 0-150Hz | Output noise should not influence the test. A value a 5µV is suitable for this requirement. Can be verified by monitoring the signal in the device under test using a “diagnostic” filter setting. | <5µV |
Time accuracy | IEC 60601-2-51 does not provide any limits. An inferred limit from the device under test. An inferred limit of ±1% is used (see 4.2). The system’s design accuracy exceeds 0.1% as a 100ppm crystal reference is used. | ±1% |
Sampling rate | A maximum sampling rate of 1kHz matches the sampling rates of ECG files. | 1kHz (8 channels) |
Power supply | Powered from the USB supply (5V 0.2A) | N/A |
Environment | Intended for normal laboratory environment. The selection of critical components such as reference voltages, DAC, precision resistors are known to be stable in the range shown. | 15-30°C 10-95% RH |